Guangzhou Rongtao Medical Tech LTD.
                                                                                                           
Verified Supplier
12 Years
Since 2014
Menu
Home Products Ultrasound Probe Repair

Ultrasound Probe Repair

China Samsung L3-12A Ultrasound Transducer Probe Strain Relief And Connector Cable Replacement for sale

Samsung L3-12A Ultrasound Transducer Probe Strain Relief And Connector Cable Replacement

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Samsung L3-12A Ultrasound Transducer Probe Strain Relief Maintenance Model: Samsung L3-12A Fault Phenomenon: The cable is wrapped with tape, the cable is aging and broken, the strain relief is aging, the image on the computer has dark shadows, and the lens is damaged. Maintenance Plan: Replace the strain relief and connector cable. Other hot models we can repair: Brand Model Aloka UST-52101 Aloka UST-52105 Aloka UST-5299 Aloka UST-5412 Aloka UST-5413 Aloka UST-9118 Aloka UST

China Aloka UST-9123 Ultrasound Transducer Probe Replace Housing Cover And Crystal for sale

Aloka UST-9123 Ultrasound Transducer Probe Replace Housing Cover And Crystal

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Aloka UST-9123 Ultrasound Transducer Probe Replace Housing Cover and Crystal Model: Aloka UST-9123 Fault Phenomenon: The connector cable is aging, the housing cover is deformed and there are gaps, the computer image is seriously missing and interference, the crystal is broken. Maintenance Plan: Replace housing cover and crystal. Other Aloka probes we can repair: Brand Model Compatible System Aloka UST-5024N-3.5 SSD-256,280 Aloka UST-52101N SSD-5500, Alpha 5, Alpha 7 Aloka ust

China GE S8-3 Ultrasound Probe Transducer Maintenance Image With Error for sale

GE S8-3 Ultrasound Probe Transducer Maintenance Image With Error

Price: Negotiable
MOQ: 1
Delivery Time: 1-5 work days

GE S8-3 Ultrasound Probe Transducer Maintenance Image with error Probe Maintenance for GE S8-3 Ultrasound Transducer Model: GE S8-3 Fault Phenomenon: There ocurred error in the test image. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain relief replacement,Electronic connector repair,Crystal/Array replacement,Plastic housing

China Mindray L14-6NE Ultrasound Transducer Probe Replace Crystal Array For Endocavity for sale

Mindray L14-6NE Ultrasound Transducer Probe Replace Crystal Array For Endocavity

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Crystal Replacement for Mindray L14-6NE Ultrasound Transducer Probe ​Model: Mindray L14-6NE Fault description: The image on the computer has dark shadow and interference, the lens bubbles and damages. Maintenance scheme: Replace crystals. Common Probe failure 1. Lens: Acoustic lens damage or bulge 2. Cable: the cable is damaged and the shell is cracked 3. Image attenuation: aging attenuation 4. Shadow: dark vertical line 5. Not recognized: probe not recognized 6. Interference

China Mindray C5-2S Ultrasound Transducer Probe Replace Crystal And Plug Connector for sale

Mindray C5-2S Ultrasound Transducer Probe Replace Crystal And Plug Connector

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Crystal and Plug Connector Replacement for Mindray C5-2S Ultrasound Transducer Probe ​Model: Mindray C5-2S Fault description: There are 21 damaged crystals in concentrated positions; The plug connector is broken or deformed. Maintenance scheme: Replace crystals and plug connector. Common Probe failure 1. Lens: Acoustic lens damage or bulge 2. Cable: the cable is damaged and the shell is cracked 3. Image attenuation: aging attenuation 4. Shadow: dark vertical line 5. Not

China L12-3 Linear Array Ultrasound Transducer Probe Replace Strain Relief Crystal And Cable for sale

L12-3 Linear Array Ultrasound Transducer Probe Replace Strain Relief Crystal And Cable

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Crystal/Cable/Strain relief Replacement for L12-3 Linear Array Ultrasound Transducer Probe ​Model: L12-3 Fault description: cable and strain relief aging, lens damage, housing slight wear, large pieces of on-machine image missing, interference, separate inspection, the crystal and the cable are damaged. Maintenance scheme: replace the strain relief, crystal and the cable. Common Probe failure 1. Lens: Acoustic lens damage or bulge 2. Cable: the cable is

China Original Ultrasound Transducer Probe Maintenance Aloka UST-9123 Connector Housing Replacement for sale

Original Ultrasound Transducer Probe Maintenance Aloka UST-9123 Connector Housing Replacement

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Connector Housing and Crystal Array Replacement for Aloka UST-9123 Ultrasound Transducer Probe ​Model: Aloka UST-9123 Fault description: The cable is aging, connector housing is deformed and has gaps. Large pieces of on-machine image are missing, interference, the crystal is damaged. Maintenance scheme: replace the crystal and connector housing. Common Probe failure 1. Lens: Acoustic lens damage or bulge 2. Cable: the cable is damaged and the shell is cracked 3. Image

China GE IC5-9-D Ultrasound Probe Repair Crystal Array Replacement for sale

GE IC5-9-D Ultrasound Probe Repair Crystal Array Replacement

Price: Negotiable
MOQ: 1
Delivery Time: 1-5 work days

GE IC5-9-D Ultrasound Probe Repair Crystal Array Replacement Model: GE IC5-9-D Fault Phenomenon: The connector cable is aging and discolored, the housing cover is worn and discolored, about 3 channels of the on-board image are missing, and the crystal is broken. Suggestion: Replace the original crystal. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range

China GE RIC5-9-D Ultrasound Probe Repair 4D Replace Dome for sale

GE RIC5-9-D Ultrasound Probe Repair 4D Replace Dome

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Ultrasound Transducer Repair Service GE RIC5-9-D Repair 4D Replace Dome Model: GE RIC5-9-D Fault Phenomenon: The dome is damaged and oil leakage, 4D doesn't work Suggestion: Repair 4D and Replace the Dome(oil filling). 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain relief replacement,Electronic connector repair,Crystal/Array

China GE RIC5-9W-RS Ultrasound Probe Repair 4D Replace Dome And Oil Filling for sale

GE RIC5-9W-RS Ultrasound Probe Repair 4D Replace Dome And Oil Filling

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Ultrasound Transducer Repair Service GE RIC5-9W-RS Replace Dome and Oil Filling Model: GE RIC5-9W-RS Fault Phenomenon: The dome is damaged and oil leakage, there are no obvious abnormalities in the image. Suggestion: Replace the Dome and oil filling. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain relief replacement

China GE 8C-RS Ultrasound Probe Repair Service Acoustic Lens Replacement for sale

GE 8C-RS Ultrasound Probe Repair Service Acoustic Lens Replacement

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Ultrasound Transducer Repair Service GE 8C-RS Acoustic Lens Replacement Model: GE 8C-RS Fault Phenomenon: The connector cable and strain relief are aging, the image on the computer has a dark shadow, and the acoustic lens is damaged. Suggestion: Replace the acoustic lens. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain relief

China GE RAB4-8L Ultrasound Transducer Probe Repair 4D Maintenance for sale

GE RAB4-8L Ultrasound Transducer Probe Repair 4D Maintenance

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Probe Repair GE RAB4-8L Ultrasound Transducer Probe 4D Maintenance GE RAB4-8L Ultrasound Transducer Probe 4D Maintenance Model: GE RAB4-8L Fault Phenomenon: 4D error, a little oil leakage, a little missing image (small impact) Suggestion: repair 4D. Suggestion: Repair the 4D. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain

China GE RIC5-9-D Ultrasound Probe Repair Crystal Array 4D Dome Cable Replacement for sale Video

GE RIC5-9-D Ultrasound Probe Repair Crystal Array 4D Dome Cable Replacement

Price: Negotiable
MOQ: 1
Delivery Time: 1-5 work days

Replace GE RIC5-9-D Crystal Array/4D/Dome/Cable Repair time: 3-5 days Warranty: 60 days after fixed We provide repair service on most common probes, 3D/4D transducers, and Endo/TEE probes. With 10 years experiences on probe repair, our technical engineer team own the most advanced probe repair processes, which enables us to fix probes from a basic lens fault to 4D /TEE fault. Once repaired, all probes will go through an extensive testing process ,until reach their optimal

China GE 4C Convex Ultrasound Probe Repair Transducer Miantenance for sale

GE 4C Convex Ultrasound Probe Repair Transducer Miantenance

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Probe Repair GE 4C Convex Probe Ultrasound Transducer Miantenance Model: GE 4C Fault Phenomenon: The cable is aging and broken, the strain relief is aging, the connector housing has wear marks, the lens is slightly damaged, the image on the machine is seriously missing, the crystal array is broken. And there has interference if shake the cable, and the image is missing, the cable is broken. Suggestion: Replace the cable, strain relief, connector housing, lens and crystal

China GE RIC5-9A-RS 4D Ultrasound Probe Repair Crystal And Dome Replace 4D Circuit for sale

GE RIC5-9A-RS 4D Ultrasound Probe Repair Crystal And Dome Replace 4D Circuit

Price: Negotiable
MOQ: 1
Delivery Time: 1-5 work days

GE RIC5-9A-RS 4D Ultrasound Probe Repair Crystal And Dome Replace 4D And Circuit Repair GE RIC5-9A-RS Repair Crystal and Dome Replace 4D and Circuit Model: GE RIC5-9A-RS Fault Phenomenon: The cable is aging and broken, the strain relief is aging, the connector housing has wear marks, the lens is slightly damaged, the image on the machine is seriously missing, the crystal array is broken. And there has interference if shake the cable, and the image is missing, the cable is

China GE RIC5-9A-RS Ultrasound Transducer 4D Repairing Crystal Array And Dome Replacement for sale

GE RIC5-9A-RS Ultrasound Transducer 4D Repairing Crystal Array And Dome Replacement

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

GE RIC5-9A-RS Ultrasound Transducer Crystal Array and Dome Replacement 4D and Circuit Maintenance Model: GE RIC5-9A-RS Fault Phenomenon: 4D error, and the image is missing. Suggestion: Replace the original crystal, replace the dome, repair the 4D. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain relief replacement,Electronic

China GE RIC6-12-D Ultrasound Probe Repair 4D Dome Replacement for sale

GE RIC6-12-D Ultrasound Probe Repair 4D Dome Replacement

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Probe Miantenance GE RIC6-12-D Repair 4D Dome Replacement GE RIC6-12-D Repair 4D and Dome Replacement Model: GE RIC6-12-D Fault Phenomenon: Error reported in the 4D, oil sac falling off and leaking oil. Suggestion: Repair 4D and Replace the Dome. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain relief replacement,Electronic

China GE RIC5-9-D Ultrasound Probe Repair 4D Replace Dome And Crystal Array for sale

GE RIC5-9-D Ultrasound Probe Repair 4D Replace Dome And Crystal Array

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Probe Miantenance GE RIC5-9-D Repair 4D Replace Dome and Crystal Array GE RIC5-9-D Repair 4D Replace Dome and Crystal Array Model: GE RIC5-9-D Fault Phenomenon: The image is seriously missing, 4D doesn't work Suggestion: Repair 4D and Replace the Dome and Crystal Array. 2D Probe Repair Service Brand GE,,Siemens,Hitachi Aloka,Toshiba,Biosound Biosound,Sumsang Medison Service Type Linear,Convex,Phased,Sector,Endocavity Repair Range Membrane/lens replacement,Strain relief

China Samsung L3-12A Ultrasound Probe Transducer Replce Strain Relief Cable And Lense for sale

Samsung L3-12A Ultrasound Probe Transducer Replce Strain Relief Cable And Lense

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Samsung L3-12A Ultrasound Probe Transducer Replace Strain Relief/Cable and Lense Model: Samsung L3-12A Fault Phenomenon: The cable is wrapped with tape, the cable is aged and broken, the strain relief is aged, the image on the machine has a dark shadow, and the lense is damaged. Maintenance Plan: Replace the strain relief, cable and lense. Other hot models we can repair: Brand Model Aloka UST-52101 Aloka UST-52105 Aloka UST-5299 Aloka UST-5412 Aloka UST-5413 Aloka UST-9118

China Samsung CA2-8AD Ultrasound Probe Transducer Replce Strain Relief And Cable for sale

Samsung CA2-8AD Ultrasound Probe Transducer Replce Strain Relief And Cable

Price: Negotiable
MOQ: 1
Delivery Time: 1-3 work days

Samsung CA2-8AD Ultrasound Probe Transducer Replace Strain Relief and Cable Model: Samsung CA2-8AD Fault Phenomenon: The cable is wrapped with tape, the cable is aged and broken, the strain relief is aged, but it can not be detected on the machine. Maintenance Plan: Replace the strain relief and cable. Other hot models we can repair: Brand Model Aloka UST-52101 Aloka UST-52105 Aloka UST-5299 Aloka UST-5412 Aloka UST-5413 Aloka UST-9118 Aloka UST-984-5 Aloka UST-9123 Aloka UST