50 Tests Colloidal Gold Swab Test Kit COVID-19 Antigen Rapid Test
COVID-19 Antigen Rapid Test Kit (Colloidal Gold)
1.【CATALOG NO. AND PACKING SPECIFICATION 】
| Catalog NO. | Packing Specification |
| BG1906C | 1 Test/Kit |
| BG1907C | 25 Tests/Kit |
| BG1908C | 50 Tests/Kit |
2.【INTENDED USE】
The COVID-19 Antigen Test Kit is used for rapid qualitative detection of SARS-CoV-2 virus nucleocapsid antigen in the anterior nasal cavity (wet nasal cavity) of the suspected sars-cov-2 patients within 7 days after symptoms.
Only for in vitro diagnostic use.【STORAGE AND STABILITY】
1. Store at 2℃ to 30℃ in the sealed pouch up to the expiration date (24 months).
2. Keep away from sunlight, moisture and heat.
3. DO NOT FREEZE.
4. When the humidity is below 60%, use it within 1 hour after opening. When the humidity is above 60%, use it immediately after opening. Production date, expiry date will be in the label.
3.【TEST PROCEDURE】
1. Specimen extraction
- Add 0.5mL (about 10 drops) of the sample extraction buffer into the extraction tube.
- Insert the swab into the extraction tube which contains 0.5mL of the extraction buffer. Roll the swab at least 6 times while pressing the head against the bottom and side of the extraction tube.
- Leave the swab in the extraction tube for 1 minute.
- Squeeze the tube several times with fingers from outside of the tube to immerse the swab. Remove the swab. The extracted solution will be used as test sample.
- Fit the dropper tip with filter on top of the extraction tube tightly.
2. Detection operations:
- Open a pouch containing a test cassette. Place the test cassette on a dry, horizontal work surface.
- Add 2 drops (about 60μl) of sample solution extract to the sample well of the test cassette.
- Observe the results showed within 10-15 minutes, and the results showed after 15 minutes have no clinical significance.
4.【INTERPRETATION OF RESULTS】
Negative : Only a red line appears in the quality control area (C), and no line appears in the test area (T).
Positive : Two red lines appear. One is in the test area (T) and the other is in the quality control area (C).
Invalid : No red line displays in the quality control area (C). This indicates that the incorrect operation or the test cassette has deteriorated or damaged. Repeat the test with a new kit. If the problem persists, stop using this lot number immediately and contact your local supplier.