Animal Care Veterinary Autoclave With Safety Door Lock and Pneumatic Seal
Animal Care Veterinary Autoclave With Safety Door Lock and Pneumatic Seal
Description
The small line of veterinary autoclaves are used in animal care facilities worldwide, incorporating the latest techniques to effectively sterilize heat and moisture-stable materials in a timely and efficient manner.
Equipped with state-of-the-art technology, these models are both reliable and easy to use. Multiple cycle options (vacuum, gravity, liquid) come standard, with a full temperature range.
Competitive Advantage
BSL3 Effluent Decontamination Sterilizers
Specification
Model | Chamber Size | Overall Size | Net Weight | Steam | Water | Power |
W*H*D mm | W*H*D mm | Kg | Kg/cycle | Kg/cycle | KVA | |
MD-0.11V | 400*400*700 | 1000*1800*1040 | 650 | 14 | 15 | 2 |
MD-0.16V | 400*400*1000 | 1000*1800*1340 | 700 | 16 | 20 | 2 |
MD-0.20V | 400*400*1300 | 1000*1800*1640 | 750 | 18 | 25 | 2 |
MD-0.24V | 600*600*670 | 1200*2000*1010 | 800 | 20 | 30 | 2 |
MD-0.30V | 600*600*840 | 1200*2000*1180 | 850 | 22 | 35 | 2 |
MD-0.36V | 600*600*1000 | 1200*2000*1340 | 900 | 25 | 40 | 2 |
MD-0.45V | 600*600*1250 | 1200*2000*1590 | 980 | 27 | 50 | 2 |
MD-0.60V | 600*600*1670 | 1200*2000*2010 | 1100 | 30 | 60 | 3 |
Water
- Softened water (Hardness <50ppm, pH Neutral) is at least requirement for the autoclaves’
durability and safety of the goods to be sterilized.
- Water pressure: 0.15----0.3MPa
Compressed air
- Oil free compressed air is important to the autoclave’s running.
- Oil free silence air-compressor is optional for user’s selection.
- Compressed air pressure: 0.5---0.8MPa
Loading patterns
- Flexible loading system suitable for various goods requirements
- Loading Baskets
- Loading trays
Electrical power
- Our ordinary electrical power is AC380, 50Hz, 3phase/N/E
- Other electrical power standards are available and compatible
Steam feeding
- Built-in electrically heated steam generator (Standard)
- External steam supply, saturated steam offer by user
- Combining steam generators: individual electrically heated steam generators
- Steam pressure: 0.4---0.7MPa
Sterilization Cycles
Cycle name | Further details | Application |
Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. | Solid materials to be sterilized: |
Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. | Materials to be sterilized from which air can’t be easily displaced |
User Programmable | All preset parameters can be reset; Multiple customized programs can be set and stored by user. | Customized |
Test cycles | ||
B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. | Test program |
Vacuum Leakage Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. | Test program |