User Friendly HMI Laboratory Autoclaves With Microcomputer With Self Diagnostic Feature

 

 

Description

 

Choosing the right steam autoclave depends on several considerations: the type of loads, frequency of use, available services and load volumes. Medpharm line of large capacity autoclaves successfully meets the challenges posed by sterilization in laboratories and research institutes.

 

Features

 

  • Counting of sterilizations performed
  • Real-time printing of cycle parameters
  • Safety-stop system for door movement in case of obstruction
  • Safety system against door opening in mid-cycle in case of power failure
  • Integral steam jacket
  • Easy to perceive and handle touch screen 
  • Computerized command of cycles
  • Chamber fine polishing
  • Chamber and jacket in AISI 304 stainless steel respectively
  • Equipment fully built in stainless steel, free of ferrous materials
  • Accurate pressure and temperature sensors
  • Optionally supplied with 1 or 2 sliding doors

 

Specifications

 

  • Model:MD-0.36V
  • Chamber Size(W*H*D mm):600*600*1000
  • Overall Size(W*H*D mm):1200*2000*1340
  • Net Weight:900kg
  • Steam:25Kg/cycle
  • Water:40Kg/cycle
  • Power:2KVA+18KW for steam generator

 

Energy Sources

 

Electrical power Our ordinary electrical power is AC380, 50Hz, 3phase, N, E
Other electrical power standards are available and compatible
Steam feeding External steam supply, 0.4---0.7MPa, saturated steam offer by user
Built-in electrically heated steam generator (Standard)
Combining steam generators: individual electrically heated steam generators or natural gas heated steam generators
Water Softened water (Hardness <50ppm, pH Neutral) is at least requirement for the autoclaves’ durability and safety of the goods to be sterilized.
Water pressure: 0.15----0.3MPa
Compressed air Oil free compressed air is important to the autoclave’s running.
Oil free silence air-compressor is optional for user’s selection.
Compressed air pressure: 0.5---0.8MPa
Loading patterns Flexible loading system suitable for various goods requirements
Loading Baskets
Loading plates(Standard)
Loading car and trolley

 

Applications

 

The autoclaves cover a wide range of applications for laboratories in research institutes, universities, pharmaceutical and biotechnological industries.

 

We Are The Difference

 

Protecting People And The Environment

 

There are many types of materials in laboratories, hygiene standards are extremely high. Perfect sterilization processes are the prerequisite for safety in laboratory operation. Suitable methods have to be used to validate sterilization processes in terms of the desired effect, because perfect sterilization results can only be achieved by precisely matching sterilization processes to the material to be sterilized.

 

Constantly High Operational Safety

 

As a result of our comprehensive service network our customers benefit from rapid reaction times,immediate availability of spare parts and personal and competent local support. Our primary goals are safety, productivity and system availability.

 

Demand-Oriented Processes

 

The diversity of sterile goods requires sterilization processes to be optimally adjusted to the respective task. Medpharm provides suitable processes and programs for every field of application. We optimize the corresponding processes subject to needs.

 

 

 

Brochure load down from here

Get in Touch

Have questions about our products or want to discuss a custom order? Our team is ready to help you.

Company ZHANGJIAGANG MEDPHARM MACHINERY LTD.
Location B631, HUAFANG INTERNATIONAL MANSION, No.178 CHENGBEI STR., ZHANGJIAGANG, JIANGSU, PRC, 215600
Contact Person Juliet

Request A Quote

Please check your email address.
Your message must be at least 20 characters.